18.118.120.204
18.118.120.204
close menu
KCI 등재 SCIE SCOPUS
용해실리카의 기준경도를 이용한 나노 압입시스템의 실험적 교정
Calibration of Nanoindentation Systems Based on the Reference Hardness of a Fused Silica
이윤희 ( Yun Hee Lee ) , 안승민 ( Seung Min Ahn ) , 김주영 ( Ju Young Kim ) , 박찬평 ( Chan Pyung Park ) , 장희광 ( Hee Kwang Jang )
UCI I410-ECN-0102-2015-500-002122492
* 발행 기관의 요청으로 무료로 이용 가능한 자료입니다.

Indirect calibration of a nanoindentation system was performed by deriving a reference loading curve. This reference load-depth relationship was accomplished by assuming a perfect nanoindentation test on an ideal reference material; if all components in a nanoindentation system are well calibrated and the reference material has unique hardness and Young’s modulus, the indentation depth can be formulated with a function of the indentation load having two parts of contact and surrounding elastic depths. Substituting both the hardness and Young’s modulus of a fused silica into the derived function, the reference loading curve was plotted and directly compared with experimental data of the fused silica measured by two commercial nanoindenters. All indentation loading curves showed leftward shifts compared to the reference loading curve, and this is discussed, offering two reasons for the result, system compliance and indenter apex geometry.(Received January 6, 2014)

[자료제공 : 네이버학술정보]
×